D/A Calibration (for PCI-1710/1710HG only)
In a way similar to the ADCAL.EXE program, the DACAL.EXE
program leads you through the whole D/A calibration procedure.
You can either use the on-board -5 V (-10 V) internal reference
voltage or use an external reference. If you use an external reference,
connect a reference voltage within the range ±10 V to the reference
input of the D/A output channel you want to calibrate. Adjust the full
scale (gain) of D/A channel 0 and 1, with VR4 and VR5 respectively.
Note:
Using a precision voltmeter to calibrate the D/A outputs is recommended.
Set the D/A data register to 4095 and adjust VR3 until the D/A
output voltage equals the reference voltage minus 1 LSB, but with
the opposite sign. For example, if V
ref
is -5 V, then V
out
should be
+4.9959 V. If V
ref
is -10 V, V
out
should be +9.9918 V.
Self A/D Calibration
Under many conditions, it is difficult to find a good enough DC
voltage source for A/D calibration. There is a simple method to solve
this problem. First, you should calibrate D/A channel 0, DA0_OUT,
with internal reference -5 V, and D/A channel 1, DA1_OUT, with
reference -10 V.